Citation: | XIAO Hong, GAO Linfeng, YAO Jie, QIAN Aijun, WANG Bin, JIANG Zhou, YUAN Weiye, BAI Jiangtao, LU Shunqi, XIU Zenghe. Effectiveness of personal dose monitoring intercomparison results in a laboratory in Shanghai[J]. Journal of Environmental and Occupational Medicine, 2023, 40(5): 571-576. DOI: 10.11836/JEOM22401 |
Individual monitoring of occupational external exposure is an essential part of the occupational health management of radiation workers, and is an important basis for the evaluation of individual absorbed dose and the diagnosis of occupational radiation diseases. Continuous participation of monitoring service providers in intercomparison is a fundamental quality assurance for routine monitoring, which can identify problems and improve them in time.
Taking the Laboratory of Radiation Protection in Shanghai Institute of Preventive Medicine as an example, to evaluate the performance of an individual occupational external dose monitoring system in the laboratory, identify influencing factors of the monitoring results, and provide a basis for improving the quality of daily monitoring by analyzing the process and results of a national intercomparison of individual dose monitoring.
According to the Testing criteria of personnel dosimetry performance for external exposure (GBZ 207-2016), and the relevant requirements of Class II (photon) inspection, a total of 20 groups of blind sample dosimeters were measured for four consecutive years from 2018 to 2021. The radiation energy source of each group was identified, and related personal dose equivalent Hp(10), the uncertainty of measurement results, and the deviation between the reported value and the reference value were calculated. The national intercomparison process and results of individual dose monitoring were also analyzed.
The energy sources of the blind samples in the tested laboratory for four years were N100 or Cs-137. The reported dose values of the blind samples were 0.57-4.61 mSv, the combined uncertainties were 0.043-0.365 mSv, the expanded uncertainties (k=2) were 0.09-0.73 mSv, and the relative expanded uncertainties (k=2) were 13.8%-16.4%. The single-group performance ∣Pi∣ of 20 sets of blind samples in the four years was ≤0.10, the yearly comprehensive performance of 5 sets of blind samples was ≤0.10, and the yearly Q score of the test report was >15 points. The laboratory achieved excellent results in the national intercomparison of individual dose monitoring in four consecutive years, except the Q value not reaching full score.
The laboratory exhibits standardized data processing of individual dose monitoring, generates accurate and reliable results, and meets the requirements of relevant national standards; but it should continue to participate in the national intercomparison of individual dose monitoring, strengthen the angular response research of energy identified dosimeter, improve the monitoring ability of low-dose X-rays, analyze the key points of reducing the uncertainty of measurement results, and continuously improve the monitoring ability.
[1] |
王恺怡, 丁艳秋, 胡爱英, 等. 2020年全国外照射个人剂量监测能力考核结果与分析[J]. 中国辐射卫生, 2021, 30(5): 555-559.
WANG K Y, DING Y Q, HU A Y, et al. Results and analysis of monitoring ability assessment of external exposure personal dose in 2020[J]. Chin J Radiol Health, 2021, 30(5): 555-559.
|
[2] |
丁艳秋, 王恺怡, 张璇, 等. 2015—2019年全国外照射个人剂量监测能力考核结果分析与总结[J]. 中华放射医学与防护杂志, 2021, 41(3): 194-198.
DING Y Q, WANG K Y, ZHANG X, et al. Summary and analysis of the intercomparison results of national individual monitoring for external exposure during 2015—2019[J]. Chin J Radiol Med Prot, 2021, 41(3): 194-198.
|
[3] |
张璇, 郭文, 丁艳秋, 等. 2017年全国外照射个人剂量监测能力考核不确定度评定分析[J]. 中华放射医学与防护杂志, 2018, 38(9): 700-704.
ZHANG X, GUO W, DING Y Q, et al. An analysis of uncertainty assessment of external individual dose monitoring in 2017 in China[J]. Chin J Radiol Med Prot, 2018, 38(9): 700-704.
|
[4] |
丁艳秋, 郭文, 胡爱英, 等. CTLD-J4000型能量鉴别式个人剂量计的剂量学实验研究[J]. 中华放射医学与防护杂志, 2018, 38(5): 374-378.
DING Y Q, GUO W, HU A Y, et al. Experimental study of dosimetry for CTLD-J4000 energy differential type personal dosimeter[J]. Chin J Radiol Med Prot, 2018, 38(5): 374-378.
|
[5] |
宋彬, 许哲, 魏召阳, 等. 2018年度苏州市个人剂量监测能力考核结果分析[J]. 中国辐射卫生, 2019, 28(4): 436-439.
SONG B, XU Z, WEI Z Y, et al. Analysis of the test results of external individual dose monitoring ability in 2018 in Suzhou[J]. Chin J Radiol Health, 2019, 28(4): 436-439.
|
[6] |
雷淑钦, 刘峰, 朱东升, 等. 新疆参加全国个人剂量监测系统能力考核结果分析[J]. 首都公共卫生, 2019, 13(4): 215-217.
LEI S Q, LIU F, ZHU D S, et al. Assessment of the results of Xinjiang's participation in the national individual dose monitoring system[J]. Cap J Public Health, 2019, 13(4): 215-217.
|
[7] |
翟贺争, 杜仲庆, 吴香君, 等. 2016—2018年参加全国个人剂量监测能力考核结果分析[J]. 中国辐射卫生, 2020, 29(1): 45-48.
ZHAI H Z, DU Z Q, WU X J, et al. Comparison and analysis of the intercomparison results of national personal dose monitoring from 2016 to 2018[J]. Chin J Radiol Health, 2020, 29(1): 45-48.
|
[8] |
高品, 丁艳秋, 王恺怡, 等. 2021年国家级个人剂量监测能力比对结果与分析[J]. 中国辐射卫生, 2022, 31(6): 651-656.
GAO P, DING Y Q, WANG K Y, et al. Results and analysis of national personal dose monitoring intercomparison in 2021[J]. Chin J Radiol Health, 2022, 31(6): 651-656.
|
[9] |
RIZK C, LONG S, OKYAR H B, et al. Results of the joint iaea/arpansa intercomparison exercise on whole body dosemeters for photons in Asia and the pacific region[J]. Radiat Prot Dosimetry, 2019, 187(4): 418-425. doi: 10.1093/rpd/ncz182
|
[10] |
RIZK C, ASKOUNIS P, OKYAR H B, et al. Uncertainty evaluation in measurement of the personal dose equivalent at nine individual monitoring services in Asia and the pacific region[J]. Radiat Prot Dosimetry, 2020, 190(2): 217-225. doi: 10.1093/rpd/ncaa093
|
[11] |
GRIMBERGEN T, FIGEL M, MCWHAN A, et al. Eurados programme of intercomparisons for individual monitoring services: seven years of development and future plans[J]. Radiat Prot Dosimetry, 2016, 170(1/4): 90-94.
|
[12] |
陈飞, 郭文, 胡爱英, 等. 2018年全国外照射个人剂量监测能力考核结果与分析[J]. 中国辐射卫生, 2020, 29(2): 178-181.
CHEN F, GUO W, HU A Y, et al. Result discussion of the nationwide ability assessment for personal external exposure dose monitoring in 2018[J]. Chin J Radiol Health, 2020, 29(2): 178-181.
|
[1] | QIAN Aijun, WEN Xuexin, GAO Linfeng, YANGFAN Qiaochu, YAO Jie, HUANG Minpeng. Intercomparison of institutions providing individual external exposure monitoring services in Shanghai from 2021 to 2023[J]. Journal of Environmental and Occupational Medicine, 2024, 41(10): 1162-1166. DOI: 10.11836/JEOM24064 |
[2] | CHEN Zhe, ZHANG Jia, LI Zhu, JIN Chenglong, SHAO Junshan, SU Yi. Quality evaluation of water quality inspection service providers in double random supervision and inspection of swimming places in Shanghai[J]. Journal of Environmental and Occupational Medicine, 2021, 38(7): 758-762. DOI: 10.13213/j.cnki.jeom.2021.21032 |
[3] | Thoracic Imaging and Occupational Diseases Standards Group, Medical Artificial Intelligence Branch, Chinese Biomedical Engineering Society. Chinese expert consensus on pneumoconiosis data labeling specifications and quality control (2020 edition)[J]. Journal of Environmental and Occupational Medicine, 2020, 37(6): 523-529. DOI: 10.13213/j.cnki.jeom.2020.20113 |
[4] | LIN Ying , ZHANG Zhen-feng , WU Ming-yang , WANG Li-min , YIN Hui , FU Wang . Tobacco Control Responsibility and Education Ability of Staff in a Medical School[J]. Journal of Environmental and Occupational Medicine, 2015, 32(12): 1154-1157. DOI: 10.13213/j.cnki.jeom.2015.15313 |
[5] | LI Wen-jing , ZHANG Jin-yan , MA Cong-xing , JIA Yu-ping , WAN Bo-yu , ZHANG Gui-bin . Monitoring and Analysis of Air Quality in a Shopping Mall in Beijing[J]. Journal of Environmental and Occupational Medicine, 2015, 32(6): 553-556,560. DOI: 10.13213/j.cnki.jeom.2015.14305 |
[6] | CAI Jing-ting . Blind Sample Measurements of Personal External Exposure Dose in a Laboratory[J]. Journal of Environmental and Occupational Medicine, 2014, 31(5): 400-402. DOI: 10.13213/j.cnki.jeom.2014.0091 |
[7] | CHEN Wei , YANG Xiao-yong , QIN Yong-chun , MA Jia-yi , CHEN Qun , FENG Zi-ya , YU Ning-le . Quality Control of a Thermoluminescence Detection System in Entrance Surface Dose Investigation[J]. Journal of Environmental and Occupational Medicine, 2013, 30(9): 696-698. |
[8] | YANG Feng-hua . Intra-laboratory Quality Control and Management of Hygienic and Sanitary Analysis Laboratory[J]. Journal of Environmental and Occupational Medicine, 2010, 27(4): 235-237. |
[9] | XIANG Ju-xiang , YANG Qi-fa , YU Jian-zhong , HU Han-qiong , LI Fei . Preliminary Exploration of Quality Control in Occupational Health Surveillance by On-site Service[J]. Journal of Environmental and Occupational Medicine, 2010, 27(2): 112-113. |
[10] | YANG Feng-hua , ZHANG Peng . On the Procedure of Quality Control to the Lead Determination in the Air of Workplace[J]. Journal of Environmental and Occupational Medicine, 2010, 27(1): 41-43. |
1. |
王井淼,秦文,谢文亮. 2017—2023年天津市某综合性三甲医院放射工作人员职业性外照射个人剂量监测结果分析. 国际放射医学核医学杂志. 2024(08): 504-508 .
![]() | |
2. |
钱爱君,温学新,高林峰,杨范翘楚,姚杰,黄敏鹏. 2021—2023年上海市外照射个人剂量监测机构能力比对结果与分析. 环境与职业医学. 2024(10): 1162-1166 .
![]() | |
3. |
杜洪凤,沈月华,游钒,秦忠雪,曹霞. 2022年四川省职业卫生技术服务机构职业卫生检测能力比对结果分析. 预防医学情报杂志. 2023(12): 1503-1507 .
![]() |