王彬, 陈晓文, 何佳伟, 秦宏冉, 白江涛, 高林峰. 核医学放射工作人员甲状腺131I内照射水平[J]. 环境与职业医学, 2023, 40(4): 462-465. DOI: 10.11836/JEOM22293
引用本文: 王彬, 陈晓文, 何佳伟, 秦宏冉, 白江涛, 高林峰. 核医学放射工作人员甲状腺131I内照射水平[J]. 环境与职业医学, 2023, 40(4): 462-465. DOI: 10.11836/JEOM22293
WANG Bin, CHEN Xiaowen, HE Jiawei, QIN Hongran, BAI Jiangtao, GAO Linfeng. Internal exposure levels of thyroid 131I in nuclear medical staff[J]. Journal of Environmental and Occupational Medicine, 2023, 40(4): 462-465. DOI: 10.11836/JEOM22293
Citation: WANG Bin, CHEN Xiaowen, HE Jiawei, QIN Hongran, BAI Jiangtao, GAO Linfeng. Internal exposure levels of thyroid 131I in nuclear medical staff[J]. Journal of Environmental and Occupational Medicine, 2023, 40(4): 462-465. DOI: 10.11836/JEOM22293

核医学放射工作人员甲状腺131I内照射水平

Internal exposure levels of thyroid 131I in nuclear medical staff

  • 摘要: 背景

    基于大量流行病学调查,放射性核素131I可以导致甲状腺疾病。

    目的

    研究上海市核医学科从事131I治疗相关放射工作人员的甲状腺131I内污染水平。

    方法

    采用调查问卷和现场监测,在普查上海市核医学基本情况的基础上,对开展131I治疗相关的放射工作人员进行内照射污染水平评估。使用便携式γ谱仪现场监测,根据测量结果估算摄入量和年待积有效剂量,对核素检出人员收集体检报告,并进行染色体检测。

    结果

    上海市核医学工作人员579人,从事131I治疗相关工作人员175名,其中18人甲状腺检出131I,检出率10.3%,其中护士和保洁人员甲状腺放射性核素131I检出率分别为23%、25%。甲状腺131I测量活度平均306.1 Bq,最大测量活度3716.9 Bq(护士)。估算摄入量最大值37544 Bq,中位数786.4 Bq。年平均待积有效剂量1.22 mSv,最高14.87 mSv。所有检出人员的染色体畸变率检测结果均正常,除一人血红蛋白偏低外,其余血象指标均正常。

    结论

    尽管本次调查的从事131I治疗放射工作人员年有限剂量未超过国家标准限值,但护士和保洁人员内照射污染不容忽视。在加强防护和去污的基础上,应开展常规内照射监测。

     

    Abstract: Background

    Based on numerous epidemiological studies, radionuclide 131I can result in thyroid disease.

    Objective

    To study the contamination level of thyroid 131I in 131I treatment associated radiation workers in nuclear medicine departments in Shanghai.

    Methods

    Based on a general survey on basic situation of nuclear medicine in Shanghai, the level of internal exposure of radiation workers involved in 131I treatment was assessed with questionnaires and on-site monitoring. A portable γ spectrometer was used for on-site detection, the intake and annual effective dose were estimated according to the measurement results. Physical examination reports were collected for radionuclide positive workers, and chromosome testing was performed.

    Results

    There were 579 nuclear medicine workers and 175 were engaged in 131I treatment in Shanghai. 131I was detected in thyroid of 18 workers, with a detection rate of 10.3%. Among them, the detection rates of thyroid 131I in nurses and cleaners were 23% and 25% respectively. The mean measured activity of thyroid 131I was 306.1 Bq and the maximum measured activity was 3716.9 Bq (nurse). The maximum intake was estimated at 37544 Bq and the median was 786.4 Bq. The average annual committed effective dose was 1.22 mSv and the maximum value was 14.87mSv. The chromosomal aberration rates of all detected workers were negative, and all blood indicators were normal except one with low hemoglobin.

    Conclusion

    The annual dose of nuclear medical staff engaged in 131I treatment does not exceed the national standard limit, but the internal exposure of nurses and cleaners should not be ignored. On the basis of strengthening protection and decontamination, routine internal exposure monitoring should be carried out.

     

/

返回文章
返回